Description:
|
dcmReset OBJECT-TYPE
SYNTAX INTEGER {other(1), hardReset(2), softReset(3), commReset(4), factoryReset(5), userReset(6), tableClear(7)}
ACCESS read-write
STATUS optional
DESCRIPTION
"indicates the different types of resets that may be issued to the dcmdevice. The values and their meaning are:
1=Other
2=Hard Reset
-(restart program, reset comms, clear all user settings, resetsensor boards, reset keyboard/display, reset cpu program memory,
does not clear data files, resets to all factory defaults, stopscurrent recording study)
3=Soft Reset
-(restart program, does not reset comms, does not change usersettings, does not reset sensor boards, resetskeyboard/display, resets cpu program memory, does not clear
data files, does not reset to factory defaults, stops current
recording study)
4=Comms Reset
-Resets the comms channel used to communicate with the field
unit to user defaults.
-It is recommended that all required elements are resynched bythe dcm device to ensure that communications is not lost with
the dcm unit (e.g. modem).
5=Factory Reset
-User settings consist of all structures that can be written to.
-Resets all user settings to factory defaults.
6=User Reset
-Which objects/tables to reset and the values to reset them to
is configured by the dcmURCTTable7=Table Clear
-The clear operation is based on the dcmTableClearTable"
|
|
|
|
|
|